BC2000-01 CO2 Analyzer

BC2000-01 analyzer

The BC2000-01 analyzer measures CO2 concentration in the atmosphere within 300÷3000 ppm. The instrument is based on the CO2 molecules absorption of the laser light emitted by the instrument's diode. Concentration is linearly related to laser beam weakening. The surveying process adopts the double tone frequency modulation technique. This technique allows an efficient reduction of the measurement noise. The sensor has been designed to be used in air but, on demand, it can be equipped with a multipass cell to augment the sensitivity. The system is wholly controlled with an internal CPU. The CPU manages the measurements, and analyzes and transmits data.

Technical data

  • Surveying: double tone frequency modulation, measurement frequency 8 MHz.
  • CPU: DX 266 MHz
  • Operating temperature: 0±40 °C (this interval can vary, depending on the laser characteristic)
  • Acquired data digital filtering

Performances

@ for one meter optical path
Concentration Interval 300÷3000 ppm
Resolution 0.2 ppm
Short term stability (15') < 1 ppm
Long term stability (1 day) <10 ppm
Signal acquisition speed 6 Hz